than 116%, analyze the precision and recovery (PAR) standard. 9.3.2.5.1 If the recovery of the PAR is outside the 78 to 116% range, the analytical system is out of control. In this case, repair the problem and repeat the analysis of the sample batch, including the MS/MSD. 9.3.2.5.2 If the recovery of the PAR is within the range of 78 to 116%, dilute the sample, MS, and MSD by a factor of 2-10 (to remain within the working range of the analytical system) and re-analyze. If the results of the dilute analyses remain outside of the acceptable range, these results may not be reported for regulatory compliance purposes. In this case, choose another sample for the MS MSD and repeat the analysis of the sample batch. 9.4 Blanks. 9.4.1 Reagent water blanks: Analyzed to demonstrate freedom from contamination. 9.4.1.1 Analyze a reagent water blank with each batch of samples. The blank must be analyzed immediately preceding calibration verification to allow for blank subtraction and to demonstrate freedom from contamination and memory effects, and must include all details of the procedure to be followed when analyzing samples. 9.4.1.2 Prepare the reagent water blank using a volume of reagent water equivalent to the volume used for sample preparation (Section 11.1). If using the micro-column procedure, adsorb the method blank using two columns, as described in Section 11. Combust the GAC from each column separately, as described in Section 11. 9.4.1.3 If the result from the blank from the batch method or the sum of the results from two columns is more than 20 ug/L, analFsis of samples is halted until the source of contamination is eliminated and a blank shows no evidence of contamination at this level. 9.4.2 Nitrate-washed GAC blanks; Analyzed daily to demonstrate that the GAC is free from contamination. 9.4.2.1 Nitrate-washed GAC blank for the batch procedure: Analyze a batch nitratewashed GAC blank by adding a scoop of dry GAC to the assembled filter apparatus containing the polycarbonate membrane and washing the GAC with the nitrate wash solution (Section 7.10) using the procedure in Section 11.2.6. 9.4.2.2 Nitrate-washed GAC blank for the column procedure: Analyze a column nitratewashed GAC blank by assembling two carbon columns in series and washing the columns with the nitrate wash solution (Section 7.10) using the procedure in Section 11.3.4.2. Analyze the GAC in each column separately. The results of the second analysis must be within 0.2 ug Ci - of the first. A difference greater than 0.2 ug Cl- indicates a lack of homogeneity in the GAC that could introduce unacceptable variability. If the difference ex ceeds this amount, the GAC should be replaced. 9.4.3 The result for the reagent water blank (Section 9.4.1) shall not exceed the result for the nitrate wash blank (Section 9.4.2.1 or 9.4.2.2) by more than 0.5 ug Cl-. 9.5 Granular activated carbon (GAC) batch testing: Each lot number or batch of activated carbon received from a supplier is tested once before use to ensure adequate quality. Use only GAC that meets the test criteria below. 9.5.1 Contamination test: Analyze a scoop of GAC. Reject carbon if the amount of OX exceeds 1 ug (25 ug Cl-/g). 9.5.2 Inorganic chloride adsorption test: Attempt to adsorb NaCl from 100 mL of a solution containing 100 mg/L in reagent water. Wash with nitrate solution and analyze. The amount of halide should be less than 1 ug Cllarger than the blank. A larger amount indicates significant uptake of inorganic chloride by the carbon. Reject carbon if the 1 ug level is exceeded. 9.6 Samples that are being used for regulatory compliance purposes shall be analyzed in duplicate. 9.6.1 The procedure for preparing duplicate sample aliquots is described in Section 11.5. 9.6.2 Calculate the RPD by following the same procedure described in Section 12.4. 9.6.3 If the RPD is greater than 20%, the analyses must be repeated. 9.6.4 If the RPD remains greater than 20%, the result may not be reported for regulatory compliance purposes. 9.7 The specifications in this method can be met if the apparatus used is calibrated properly and maintained in a calibrated state. The standards used for calibration (Section 10), calibration verification (Section 9.9), and for initial (Section 9.2.2) and ongoing (Section 9.10) precision and recovery should be identical, so that the most precise results will be obtained. 9.8 Depending on specific program quirements, field duplicates may be collected to determine the precision of the sampling technique. 9.9 At the beginning and end of each eight-hour shift during which analyses are performed, system performance and calibration are verified. Verification of system performance and calibration may be performed more frequently, if desired. 9.9.1 If performance and calibration are verified at the beginning and end of each shift (or more frequently), samples analyzed during that period are considered valid. 9.9.2 If performance and calibration are not verified at both the beginning and end of a shift (or more frequently), samples analyzed during that period must be reanalyzed. 9.9.3 If calibration is verified at the beginning of a shift, recalibration using the five standards described in Section 10.6 is not re necessary; otherwise, the instrument must be recalibrated prior to analyzing samples (Section 10). 9.9.4 Cell maintenance and other changes to the analytical system that can affect system performance may not be performed during the eight-hour (or shorter) shift. 9.10 Calibration verification and ongoing precision and recovery: Calibration and system performance are verified by the analysis of the 100 ug/L PAR standard. 9.10.1 Analyze a blank (Section 9.4) and analyze the PAR standard (Section 7.12.4) immediately thereafter at the beginning and end of each shift. Compute the concentration of organic halide in the blank and in the PAR standard using the procedures in Section 12. The blank shall be less than 2 ug Cl(20 ug/L equivalent). 9.10.2 Subtract the result for the blank from the result of the PAR standard using the procedures in Section 12, and compute the percent recovery of the blank-subtracted PAR standard. The percent recovery shall be in the range of 78 to 116%. 9.10.3 If the recovery is within this range, the analytical process is in control and analysis of blanks and samples may proceed. If, however, the recovery is not within the acceptable range, the analytical process is not in control. In this event, correct the problem and repeat the ongoing precision and recovery test (Section 9.10), or recalibrate (Sections 10.5 through 10.6). 9.10.4 If the recovery is not within the acceptable range for the PAR standard analyzed at the end of the eight-hour shift, correct the problem, repeat the ongoing precision and recovery test (Section 9.10), or recalibrate (Sections 10.5 through 10.6), and reanalyze the sample batch that was analyzed during the eight-hour shift. 9.10.5 If the recovery is within the acceptable range at the end of the shift, and samples are to be analyzed during the next eighthour shift, the end of shift verification may be used as the beginning of shift verification for the subsequent shift, provided the next eight-hour shift begins as the first shift ends. 9.11 It is suggested but not required that the laboratory develop a statement of data quality for AOX and develop QC charts to form a graphic demonstration of method performance. Add results that pass the specification in Section 9.10.2 to initial and previous ongoing data. Develop a statement of data quality by calculating the average percent recovery (R) and the standard deviation of percent recovery (sr). Express the accuracy as a recovery interval from R – 2s, to R = 28,. For example, if R=95% and s=5%, the accuracy is 85 to 105%. and models of instruments will require different operating procedures. Laboratories should follow the operating instructions provided by the manufacturer of their particular instrument. Sensitivity, instrument detection limit, precision, linear range, and interference effects must be investigated and established for each particular instrument. Calibration is performed when the instrument is first set up and when calibration cannot be verified (Section 9.9). 10.2 Cell performance test: Inject 100 uL of the sodium chloride solution (10 ug Cl-; Section 7.4) directly into the titration cell electrolyte. Adjust the instrument to produce a reading of 10 ug Cl-. 10.3 Combustion system test: This test can be used to assure that the combustion micro-coulometer systems are performing properly without introduction of carbon. This test should be used during initial instrument setup and when instrument performance indicates a problem with the combustion system. 10.3.1 Designate a quartz boat for use with the ammonium chloride (NHCl) solution only. 10.3.2 Inject 100 uL of the NHCl solution (Section 7.5) into this boat and proceed with the analysis. 10.3.3 The result shall be between 9.5 and 10.5 ug Cl-. If the recovery is not between these limits, the combustion or micro-coulometer systems are not performing properly. Check the temperature of the combustion system, verify that there are no leaks in the combustion system, confirm that the cell is performing properly (Section 10.2), and then repeat the test. 10.4 Trichlorophenol combustion test: This test can be used to assure that the combustion/micro-coulometer systems are performing properly when carbon is introduced. It should be used during instrument setup and when it is necessary to isolate the adsorption and combustion steps. 10.4.1 Inject 10 uL of the 1 mg/mL trichlorophenol stock solution (Section 7.12.2) onto one level scoop of GAC in a quartz boat. 10.4.2 Immediately proceed with the analysis to prevent loss of trichlorophenol and to prevent contamination of the carbon. 10.4.3 The result shall be between 9.0 and 11.0 ug Cl-. If the recovery is not between these limits, the combustion/micro-coulometer system shall be adjusted and the test repeated until the result falls within these limits. 10.5 Background level of Cl-: Determine the average background level of Cl - for the entire analytical system as follows: 10.5.1 Using the procedure in Section 11 (batch or column) that will be used for the analysis of samples, determine the background level of Cl- in each of three portions of reagent water. The volume of reagent 10.0 Calibration and Standardization 10.1 Assemble the Ox system and establish the operating conditions necessary for analysis. Differences between various makes system and/or preparation of fresh calibration standards. 10.6.5 Using the average calibration factor, compute the percent recovery at each calibration point. The recovery at each calibration point shall be within the range of 80 to 111%. If any point is not within this range, a fresh calibration standard shall be prepared for that point, this standard shall be analyzed, and the calibration factor (Section 10.6.3) and calibration linearity (Section 10.6.4) shall be computed using the new calibration point. All points used in the calibration must meet the 80 to 111% recovery specification. water used shall be the same as the volume used for analysis of samples. 10.5.2 Calculate the average (mean) concentration of Cl- and the standard deviation of the concentration. 10.5.3 The sum of the average concentration plus two times the standard deviation of the concentration shall be less than 20 ug/L. If not, the water or carbon shall be replaced, or the adsorption system moved to an area free of organic halide vapors, and the test Section 10.5) shall be repeated. Only after this test is passed may calibration proceed. 10.6 Calibration by external standard: A calibration line encompassing the calibration range is developed using solutions of 2,4,6-trichlorophenol. 10.6.1 Analyze each of the five calibration solutions (Section 7.12.3) using the procedure in Section 11 (batch or column) that will be used for the analysis of samples, and the same procedure that was used for determination of the system background (Section 10.5). Analyze these solutions beginning with the lowest concentration and proceeding to the highest. Record the response of the microcoulometer to each calibration solution. 10.6.2 Prepare a method blank as described in Section 9.4. Subtract the value of the blank from each of the five calibration results, as described in Section 12. 10.6.3 Calibration factor (ratio of response to concentration) Using the blank subtracted results, compute the calibration factor at each calibration point, and compute the average calibration factor and the relative standard deviation (coefficient of variation; Cv) of the calibration factor over the calibration range. 10.6.4 Linearity: The Cy of the calibration factor shall be less than 20%; otherwise, the calibration shall be repeated after adjustment of the combustion/micro-coulometer 11.0 Procedure 11.1 Sample dilution: Many samples will contain high concentrations of halide. If analyzed without dilution, the micro-coulometer can be overloaded, resulting in frequent cell cleaning and downtime. The following guidance is provided to assist in estimating dilution levels. 11.1.1 Paper and pulp mills that employ chlorine bleaching: Samples from pulp mills that use a chlorine bleaching process may overload the micro-coulometer. To prevent system overload, the maximum volume suggested for paper industry samples that employ halide in the bleaching process is 100 mL. An adsorption volume as small as 25 mL may be used, provided the concentration of AOX in the sample can be measured reliably, as defined by the requirements in Section 9.11. To minimize volumetric error, an adsorption volume less than 25 mL may not be used. If AOX cannot be measured reliably in a 100-mL sample volume, a sample volume to a maximum of 1000 mL must be used. The sample and adsorption volumes are suggested for paper industry samples employing chlorine compounds in the bleaching process: Paper or pulp mill stream Sample volume (mL)* Adsorption volume (mL) Evaporator condensate 100.0 100 Process water 100.0 100 Pulp mill effluent 30.0 50 Paper mill effluent 10.0 25 Combined mill effluent 5.0 25 Combined bleach effluent 1.0 25 C-stage filtrate 0.5 25 E-stage filtrate 0.5 25 Assumes dilution to final volume of 100 ml. All sample aliquots (replicates, diluted samples) must be analyzed using the same fixed final volume (sample volume plus reagent water, as needed). re 11.1.2 Sample dilution procedure. 11.1.2.1 Partially fill a precleaned volumetric flask with pH < 2 reagent water, allowing for the volume of sample to be added. 11.1.2.2 Mix sample thoroughly by tumbling or shaking vigorously. 11.1.2.3 Immediately withdraw the quired sample aliquot using a pipet or microsyringe. NOTE: Because it will be necessary to rinse the pipet or micro-syringe (Section 11.1.2.5), it may be necessary to pre-calibrate the 208-169 D-8 pipet or micro-syringe to assure that the exact volume desired will be delivered. 11.1.2.4 Dispense or inject the aliquot into the volumetric flask. 11.1.2.5 Rinse the pipet or syringe with small portions of reagent water and add to the flask. 11.1.2.6 Dilute to the mark with pH < 2 reagent water. 11.1.3 All samples to be reported for regulatory compliance monitoring purposes must be analyzed in duplicate, as described in Section 11.5. 11.1.4 Pulp and Paper in-process samples: The concentration of organic halide in inprocess samples has been shown to be 20 to 30% greater using the micro-column adsorption technique than using the batch adsorption technique. For this reason, the microcolumn technique shall be used for monitoring in-process samples. Examples of inprocess samples include: combined bleach plant effluent, C-stage filtrate, and E-stage filtrate. 11.2 Batch adsorption and filtration. 11.2.1 Place the appropriate volume of sample (diluted if necessary), preserved as described in Section 8, into an Erlenmeyer flask. 11.2.2 Add 5 mL of nitrate stock solution to the sample aliquot. 11.2.3 Add one level scoop of activated carbon that has passed the quality control tests in Section 9. 11.2.4 Shake the suspension for at least one hour in a mechanical shaker. 11.2.5 Filter the suspension through a polycarbonate membrane filter. Filter by suction until the liquid level reaches the top of the carbon. 11.2.6 Wash the inside surface of the filter funnel with 25 mL (15 mL) of nitrate wash solution in several portions. After the level of the final wash reaches the top of the GAC, filter by suction until the cake is barely dry. The time required for drying should be minimized to prevent exposure of the GAC to halogen vapors in the air, but should be sufficient to permit drying of the cake so that excess water is not introduced into the combustion apparatus. A drying time of approximately 10 seconds under vacuum has been shown to be effective for this operation. 11.2.7 Carefully remove the top of the filter holder, making sure that no carbon is lost. This operation is most successfully performed by removing the clamp, tilting the top of the filter holder (the funnel portion) to one side, and lifting upward. 11.2.8 Using a squeeze bottle or micro-syringe, rapidly rinse the carbon from the inside of the filter holder onto the filter cake using small portions of wash solution. Allow the cake to dry under vacuum for no more than 10 seconds after the final rinse. Immediately turn the vacuum off. 11.2.9 Using tweezers, carefully fold the polycarbonate filter in half, then in fourths, making sure that no carbon is lost. 11.3 Column adsorption. 11.3.1 Column preparation: Prepare a sufficient number of columns for one day's operation as follows: 11.3.1.1 In a glove box or area free from halide vapors, place a plug of Cerafelt into the end of a clean glass column. 11.3.1.2 Fill the glass column with one level scoop (approximately 40 mg) of granular activated carbon that has passed the quality control tests in Section 9. 11.3.1.3 Insert a Cerafelt plug into the open end of the column to hold the carbon in place. 11.3.1.4 Store the columns in a glass jar with PTFE lined screw-cap to prevent infiltration of halide vapors from the air. 11.3.2 Column setup. 11.3.2.1 Install two columns in series in the adsorption module. 11.3.2.2 If the sample is known or expected to contain particulates that could prevent free flow of sample through the micro-columns, a Cerafelt plug is placed in the tubing ahead of the columns. If a measurement of the OX content of the particulates is desired, the Cerafelt plug can be washed with nitrate solution, placed in a combustion boat, and processed as a separate sample. 11.3.3 Adjusting sample flow rate: Because the flow rate used to load the sample onto the columns can affect the ability of the GAC to adsorb organic halides, the flow rate of the method blank is measured, and the gas pressure used to process samples is adjusted accordingly. The flow rate of the blank, which is composed of acidified reagent water and contains no particulate matter, should be greater than the flow rate of any sample containing even small amounts of particulate matter. 11.3.3.1 Fill the sample reservoir with the volume of reagent water chosen for the analysis (Section 9.4.1.2) that has been preserved and acidified as described in Section 8. Cap the reservoir. 11.3.3.2 Adjust the gas pressure per the manufacturer's instructions. Record the time required for the entire volume of reagent water to pass through both columns. The flow rate must not exceed 3 mL/min over the duration of the time required to adsorb the volume. If this flow rate is exceeded, adjust gas pressure, prepare another blank, and repeat the adsorption. 11.3.3.3 Once the flow rate for the blank has been established, the same adsorption conditions must be applied to all subsequent samples during that eight-hour shift, or until another method blank is processed, whichever comes first. To aid in overcoming breakthrough problems, a lower gas pressure (and, therefore, flow rate) may be used for or processing of samples, if desired. If the sample adsorption unit is disassembled cleaned, the flow rate must be checked before processing additional samples. 11.3.3.4 Elute the pair of columns with 2 mL of nitrate wash solution. The flow rate of nitrate wash solution must not exceed 3 mL/ min. 11.3.3.5 Separate the columns and mark for subsequent analysis. 11.3.4 The adsorption of sample volumes is performed in a similar fashion. Fill the sample reservoir with the sample volume chosen for the analysis (Section 11.1), that has been preserved as described in Section 8. All analyses must be performed with this volume (sample volume plus reagent water, as needed) in order to maintain a flow rate no greater than that determined for the blank (see Section 11.3.3). 11.3.4.1 Use the same gas pressure for sample adsorption as is used for the blank. 11.3.4.2 Elute the columns with 2 mL of the nitrate wash solution. 11.3.4.3 Separate the columns and mark for subsequent analysis. 11.3.5 If it is desirable to make measurements at levels lower than can be achieved with the sample volume chosen, or if the instrument response of an undiluted sample is less than three times the instrument response of the blank (Section 12.6.3), a larger sample volume must be used. 11.4 Combustion and titration. 11.4.1 Polycarbonate filter and GAC from batch adsorption. 11.4.1.1 Place the folded polycarbonate filter containing the GAC in a quartz combustion boat, close the airlock, and proceed with the automated sequence. 11.4.1.2 Record the signal from the microcoulometer for a minimum integration time of 10 minutes and determine the concentration of Cl- from calibration data, per Section 12. 11.4.2 Columns from column adsorption. 11.4.2.1 Using the push rod, push the carbon and the Cerafelt plug(8) from the first column into a combustion boat. Proceed with the automated sequence. 11.4.2.2 Record the signal from the microcoulometer for a minimum integration time of 10 minutes and determine the concentration of Cl - for the first column from calibration data, per Section 12. 11.4.2.3 Repeat the automated sequence with the second column. 11.4.2.4 Determine the extent of breakthrough of organic halides from the first column to the second column, as described in Section 12. 11.4.3 The two columns that are used for the method blank must be combusted separately, as is done for samples. 11.5 Duplicate sample analysis: All samples to be reported for regulatory compliance purposes must be analyzed in duplicate. This requirement ap plies to both the batch and column adsorption procedures. In addition, if it is necessary to dilute the sample for the purposes of reducing breakthrough or maintaining the concentration within the calibration range, a more or less dilute sample must be analyzed. The adsorption volumes used for analysis of undiluted samples, diluted samples, and all replicates must be the same as the volume used for QC tests and calibration (Sections 9 and 10). 11.5.1 Using results from analysis of one sample volume (Section 11.4) and the procedure in Section 11.1.2, determine if the dilution used was within the calibration range of the instrument and/or if breakthrough exceeded the specification in Section 12.3.1. If the breakthrough criterion was exceeded or the sample was not within the calibration range, adjust the dilution volume as needed. If the breakthrough criterion was not exceeded and the sample dilution was within the calibration range, a second volume at the same dilution level may be used. 11.5.2 Adsorb the sample using the same technique (batch or column) used for the first sample volume. Combust the GAC from the second volume as described in Section 11.4, and calculate the results as described in Section 12. Compare the results of the two analyses as described in Section 12.4. 11.5.3 Duplicate analyses are not required for method blanks, as different dilution leyels are not possible. 11.5.4 Duplicate analyses of the PAR standard used for calibration verification (Section 9.10) are not required. 12.0 Data Analysis and Calculations 12.1 Batch Adsorption Method: Calculate the blank-subtracted concentration of adsorbable organic halide detected in each sample (in micrograms of chloride per liter) using the following equation: (C-B) AOX(ug/L)= V Where: C=ug Cl- from micro-coulometer for the sample Brug Cl- from micro-coulometer for the rea gent water blank (Section 9.4.1) V = volume of sample in liters This calculation is performed for each of the two dilution levels analyzed for each sample. 12.2 Column Adsorption Method: Calculate the blank-subtracted concentration of adsorbable organic halide detected in each sample (in micrograms of chloride per liter) using the following equation: |
